Key numbers from this manual
As printed in the manual, with metric and US units side by side. The typical range comes from all refrigerator manuals on this site that print the same figure.
| Fridge adjustment range | -55͠ to -85͠ (1͠ increment)temperature control setting-55㷄 to -85㷄Chamber temperature settingTypical for refrigerators: 36 to 46 °F (2 to 8 °C) | p. 5 p. 16 |
|---|
Panasonic MDF-C8V1 questions
Answers taken from this manual. Each one links to the page it comes from.
What temperature can I set on the unit?
-55 to -85 ͠, adjustable in 1 ͠ increments.
Manual: page 5
What high and low temperature alarm settings are available?
The alarm ranges and initial values are:
| Alarm | Selectable range | Initial value |
|---|---|---|
| High temp. | +5 to +20 ͠ | +10 ͠ |
| Low temp. | -5 to -20 ͠ | -10 ͠ |
Manual: page 5
What are the refrigerator’s external dimensions?
W 550 × D 685 (+83)* × H 945 mm, with the asterisk marking the projected area dimension.
Manual: page 4
What happens when power returns after an outage?
If power returns within 3 seconds, the unit restarts with initial settings and the remote alarm output turns off.
Manual: page 18
How do I reset the battery accumulation time?
Use the battery accumulation reset procedure to return the stored time to ’00.0’.
- Open function F06.
- Enter ‘409’, then press SET.
- Confirm the stored time changes to ’00.0’ and the BATTERY lamp turns off.
Manual: page 18
What do the MDF-C8V1 error codes mean?
The displayed code identifies the detected sensor, battery, compressor-temperature, or fan problem.
| Code | Meaning |
|---|---|
| E01 or E02 | Temperature sensor disconnected or short-circuited |
| E05 or E06 | Compressor sensor disconnected or short-circuited |
| E09 | Battery switch is OFF |
| E10 | Abnormal compressor temperature |
| E15 | Fan motor malfunction |
Manual: page 17
What is CN3 used for?
It provides a remote alarm contact output: #1 COM., #2 N.O., and #3 N.C.
Manual: page 15
What does CN8 do during a power failure?
It connects the battery to supply the Main PCB during a power failure; #2:6V is listed.
Manual: page 15
Why does the compressor wait after power is supplied?
3 min. initially: the Compressor delay time starts the compressor after power is supplied, and a cycle stop forces the next 3 min. off.
Manual: page 22
Why do the buzzer and remote alarm sound later?
10min. initially: during a high/low temperature alarm, the buzzer and remote alarm are emitted after the delay, while the ALARM lamp and display blink without delay.
Manual: page 22
Error codes in this manual
| Code | Meaning | What to do | |
|---|---|---|---|
| E01 | Temperature sensor is disconnected. | Press BUZZER to stop the intermittent buzzer. | p. 17 |
| E02 | Temperature sensor is short circuited. | Press BUZZER to stop the intermittent buzzer. | p. 17 |
| E05 | Compressor sensor is disconnected. | Press BUZZER to stop the intermittent buzzer. | p. 17 |
| E06 | Compressor sensor is short circuited. | Press BUZZER to stop the intermittent buzzer. | p. 17 |
| E09 | Battery switch is off. | Press ALARM TEST to display E09. | p. 17 |
| E10 | Compressor temperature is abnormal. | Press BUZZER to stop the buzzer. | p. 17 |
| E15 | Fan motor is malfunctioning. | Press BUZZER to stop the buzzer. | p. 17 |
Similar Panasonic refrigerators
| Panasonic MPR-215F | Operating Instructions | 46 pages |
|---|---|---|
| Panasonic MPR-721 | Operating Instructions | 40 pages |
| Panasonic NR-XY680Y | Operating Instructions | 104 pages |
| Panasonic NR-F603GT | Operating Instructions | 80 pages |
| Panasonic Click here to download the manual | Instruction Manual | 68 pages |
| Panasonic NIR-C371 | Operating Instructions | 56 pages |
| Panasonic NR-BW530H | Operating Instructions | 56 pages |
| Panasonic NR-BX421 | Operating Instructions | 36 pages |