Sanyo error codes
20 Sanyo error and alarm codes from 11 manuals: what each code means and what the manual says to do.
20 codes11 manuals
Codes and what they mean
As described in each manual. The same code can mean different things on different models, so check the model column.
| Code | Meaning | What to do | Manual |
|---|---|---|---|
| dF | Forced defrost cycle is operating. | Wait until forced defrost completes; normal operation then resumes. | MPR-411F |
| E01 | Thermal sensor is disconnected or short-circuited. | MDF-593C | |
| E01 | Thermal sensor is short circuited. | MDF-136, MPR-1411 | |
| E01 | Thermal sensor is disconnected or short circuited. | Unit operates using the defrost sensor. | MPR-1013 |
| E01 | Thermal sensor is disconnected. | MPR-514 | |
| E02 | Thermal sensor is disconnected. | MDF-136, MPR-1411 | |
| E02 | Defrost sensor 1 is disconnected or short circuited. | Unit keeps running. | MPR-1013 |
| E02 | Thermal sensor is short circuited. | MPR-514 | |
| E03 | Cascade sensor is disconnected. | MDF-593C | |
| E03 | Battery switch is off during alarm testing. | MDF-136 | |
| E03 | Defrost sensor 2 is disconnected or short circuited. | Unit keeps running. | MPR-1013 |
| E03 | Defrost sensor is disconnected. | MPR-1411, MPR-514 | |
| E04 | Cascade sensor is short-circuited. | MDF-593C | |
| E04 | Compressor protective sensor is disconnected or short circuited. | MDF-136 | |
| E04 | Defrost sensor is short circuited. | MPR-1411, MPR-514 | |
| E05 | Filter sensor is disconnected. | MDF-593C | |
| E05 | Compressor sensor is disconnected. | MPR-1411 | |
| E06 | Filter sensor is short-circuited. | MDF-593C | |
| E06 | Compressor sensor is short circuited. | MPR-1411 | |
| E09 | Compressor cooling fan motor has failed. | MDF-136 | |
| E09 | Battery switch was turned on before required setup. | Turn off the battery switch and perform the setting procedure. | MPR-1411 |
| E09 | Battery switch was turned on before completing the required setting. | Turn off the battery switch, complete the setting, then turn it on again. | MPR-514 |
| E1 | Thermal sensor is disconnected. | MPR-311 | |
| E1 | Refrigerator thermal sensor is open or shorted. | MPR-411F | |
| E1 | Upper temperature sensor is open or short circuited. | MBR-1404G | |
| E1 | Upper monitor bottle temperature sensor is disconnected or short circuited. | MBR-304D, MBR-304 | |
| E2 | Thermal sensor is short circuited. | MPR-311 | |
| E2 | Freezer thermal sensor is open or shorted. | MPR-411F | |
| E2 | Lower temperature sensor is open or short circuited. | MBR-1404G | |
| E2 | Lower monitor bottle temperature sensor is disconnected or short circuited. | MBR-304D, MBR-304 | |
| E3 | Defrost sensor is disconnected. | MPR-311 | |
| E3 | Defrost sensor is open or shorted. | MPR-411F | |
| E3 | Defrost sensor is open or short circuited. | MBR-1404G | |
| E3 | Defrost sensor is disconnected or short circuited. | MBR-304D, MBR-304 | |
| E4 | Defrost sensor is short circuited. | MPR-311 | |
| E4 | Over temperature sensor is open or short circuited. | MBR-1404G | |
| E4 | Over temperature sensor is disconnected or short circuited. | MBR-304D, MBR-304 | |
| E10 | Compressor temperature is abnormally high. | MDF-136 | |
| E10 | Cooling compressor fan motor has failed. | MPR-1411 | |
| F-1 | Battery replacement is due after approximately three years. | Replace the battery. | MPR-1411 |
| F-2 | Fan motor replacement is due after approximately six years. | Replace the fan motor. | MPR-1411 |
| F01 | Refrigerator high temperature alarm setting | Set the first digit to 1, adjust the alarm value, and press the set key. | MPR-414F |
| F02 | Refrigerator low temperature alarm setting | Set the first digit to 2, adjust the alarm value, and press the set key. | MPR-414F |
| F03 | Freezer high temperature alarm setting | Set the first digit to 3, adjust the alarm value, and press the set key. | MPR-414F |
| F04 | Freezer low temperature alarm setting | Set the first digit to 4, adjust the alarm value, and press the set key. | MPR-414F |
| L1 | Key lock is enabled. | MPR-411F |