Codes and what they mean

As described in each manual. The same code can mean different things on different models, so check the model column.

CodeMeaningWhat to doManual
dFForced defrost cycle is operating.Wait until forced defrost completes; normal operation then resumes.MPR-411F
E01Thermal sensor is disconnected or short-circuited.MDF-593C
E01Thermal sensor is short circuited.MDF-136, MPR-1411
E01Thermal sensor is disconnected or short circuited.Unit operates using the defrost sensor.MPR-1013
E01Thermal sensor is disconnected.MPR-514
E02Thermal sensor is disconnected.MDF-136, MPR-1411
E02Defrost sensor 1 is disconnected or short circuited.Unit keeps running.MPR-1013
E02Thermal sensor is short circuited.MPR-514
E03Cascade sensor is disconnected.MDF-593C
E03Battery switch is off during alarm testing.MDF-136
E03Defrost sensor 2 is disconnected or short circuited.Unit keeps running.MPR-1013
E03Defrost sensor is disconnected.MPR-1411, MPR-514
E04Cascade sensor is short-circuited.MDF-593C
E04Compressor protective sensor is disconnected or short circuited.MDF-136
E04Defrost sensor is short circuited.MPR-1411, MPR-514
E05Filter sensor is disconnected.MDF-593C
E05Compressor sensor is disconnected.MPR-1411
E06Filter sensor is short-circuited.MDF-593C
E06Compressor sensor is short circuited.MPR-1411
E09Compressor cooling fan motor has failed.MDF-136
E09Battery switch was turned on before required setup.Turn off the battery switch and perform the setting procedure.MPR-1411
E09Battery switch was turned on before completing the required setting.Turn off the battery switch, complete the setting, then turn it on again.MPR-514
E1Thermal sensor is disconnected.MPR-311
E1Refrigerator thermal sensor is open or shorted.MPR-411F
E1Upper temperature sensor is open or short circuited.MBR-1404G
E1Upper monitor bottle temperature sensor is disconnected or short circuited.MBR-304D, MBR-304
E2Thermal sensor is short circuited.MPR-311
E2Freezer thermal sensor is open or shorted.MPR-411F
E2Lower temperature sensor is open or short circuited.MBR-1404G
E2Lower monitor bottle temperature sensor is disconnected or short circuited.MBR-304D, MBR-304
E3Defrost sensor is disconnected.MPR-311
E3Defrost sensor is open or shorted.MPR-411F
E3Defrost sensor is open or short circuited.MBR-1404G
E3Defrost sensor is disconnected or short circuited.MBR-304D, MBR-304
E4Defrost sensor is short circuited.MPR-311
E4Over temperature sensor is open or short circuited.MBR-1404G
E4Over temperature sensor is disconnected or short circuited.MBR-304D, MBR-304
E10Compressor temperature is abnormally high.MDF-136
E10Cooling compressor fan motor has failed.MPR-1411
F-1Battery replacement is due after approximately three years.Replace the battery.MPR-1411
F-2Fan motor replacement is due after approximately six years.Replace the fan motor.MPR-1411
F01Refrigerator high temperature alarm settingSet the first digit to 1, adjust the alarm value, and press the set key.MPR-414F
F02Refrigerator low temperature alarm settingSet the first digit to 2, adjust the alarm value, and press the set key.MPR-414F
F03Freezer high temperature alarm settingSet the first digit to 3, adjust the alarm value, and press the set key.MPR-414F
F04Freezer low temperature alarm settingSet the first digit to 4, adjust the alarm value, and press the set key.MPR-414F
L1Key lock is enabled.MPR-411F